A thermocouple is a device made by two different wires joined at one end, called junction end or measuring end. The two wires are called thermoelements
or legs of the thermocouple: the two thermoelements are distinguished as
positive and negative ones. The other end of the thermocouple is called
tail end or reference end. The junction end is immersed in the environment whose temperature T2
has to be measured, which can be for instance the temperature of a
furnace at about 500°C, while the tail end is held at a different
temperature T1, e.g. at ambient temperature.
Figure 1.
Because of the temperature difference between
junction end and tail end a voltage difference can be measured between
the two thermoelements at the tail end: so the thermocouple is a
temperature-voltage transducer.
The temperature vs voltage relationship is given by:
Equation 1
Where Emf is the Electro-Motive Force or Voltage produced by the thermocople at the tail end, T1 and T2 are the temperatures of reference and measuring end respectively, S12 is called Seebeck coefficient of the thermocouple and S1 and S2 are the Seebeck coefficient of the two thermoelements; the Seebeck coefficient depends on the material the thermoelement is made of. Looking at Equation1 it can be noticed that:
- a null voltage is measured if the two thermoelements are made of the same materials: different materials are needed to make a temperature sensing device,
- a null voltage is measured if no temperature difference exists between the tail end and the junction end: a temperature difference is needed to operate the thermocouple,
- the Seebeck coefficient is temperature dependent.
In order to clarify the first point let us consider the following example (Figure2):
when a temperature difference is applied between the two ends of a
single Ni wire a voltage drop is developed across the wire itself.
The end of the wire at the highest temperature, T2, is called hot end, while the end at the lowest temperature, T1, is called cold end.
Figure2: Emf produced by a single wire |
When a voltmeter, with Cu
connection wires, is used to measure the voltage drop across the Ni
wire, two junctions need to be made at the hot and cold ends between the
Cu wire and the Ni wire; assuming that the voltmeter is at room
temperature T1, one of the Cu wires of the voltmeter will
experience along it the same temperature drop from T2 to T1 the Ni wire
is experiencing. In the attempt to measure the voltage drop on the Ni
wire a Ni-Cu thermocouple has been made and so the measured voltage is
in reality the voltage drop along the Ni wire plus the voltage drop
along the Cu wire.
The Emf along a single thermoelement cannot be measured: the Emf measured at the tail end in Figure1
is the sum of the voltage drop along each of the thermoelements. As two
thermoelements are needed, the temperature measurement with
thermocuoples is a differential measurement.
Note: if the wire in Figure2 was a Cu wire a null voltage would have been measured at the voltmeter.
The temperature measurement with thermocouples is also a differential measurement because two different temperatures, T1 and T2, are involved. The desired temperature is the one at the junction end, T2. In order to have a useful transducer for measurement, a monotonic Emf versus junction end temperature T2 relationship is needed, so that for each temperature at the junction end a unique voltage is produced at the tail end.
However, from the integral in Equation1 it can be understood that the Emf depends on both T1 and T2: as T1 and T2 can change indipendently, a monotonic Emf vs T2
relationship cannot be defined if the tail end temperature is not
constant. For this reason the tail end is mantained in an ice bath made
by crushed ice and water in a Dewar flask: this produces a reference
temperature of 0°C. All the voltage versus temperature relationships for
thermocouples are referenced to 0°C.
The resulting measuring system required for a thermocople is shown in Figure3.
Figure3: A measuring system for thermocouples |
In order to measure the voltage at
the tail end, two copper wires are connected between the thermoelements
and the voltmeter: both the Cu wires experience the same temperature
difference and as a result the voltage drops along each of them are
equal to each other and cancel out in the measurement at the voltmeter.
The ice bath is usually replaced in industrial application with an integrated circuit called cold junction compensator:
in this case the tail end is at ambient temperature and the
temperature fluctuations at the tail end are tolerated; in fact the
cold junction compensator produces a voltage equal to the thermocouple
voltage between 0°C and ambient temperature, which can be added to the
voltage of the thermocouple at the tail end to reproduce the voltage
versus temperature relationship of the thermocouple.
A sketch of a thermocouple with cold junction compensation is reported in Figure4.
Figure4: An example of Cold Junction Compensation |
It should be underlined that the
cold junction compensation cannot reproduce exactly the voltage versus
temperature relationship of the thermocouple, but can only approximate
it: for this reason the cold junction compensation introduces an error
in the temperature measurement.
Figure4
shows also the filtering and amplification of the thermocouple. Being
the thermocouple voltage a DC signal, removal of AC noise through
filtering is beneficial; furthermore the thermocouples produce voltage
of few tens of mV and for this reason amplification is required. The
small voltage range for some of the most common thermocouples (letter designated thermocouples) is shown in Figure5, where their voltage versus temperature relationship is reported.
Type R, S and B thermocouples use
Pt-base thermoelements and they can operate at temperatures up to
1700°C; however they are more expensive and their voltage output is
lower than type K and type N thermocouples, which use Ni-base
thermoelements. However, Ni base thermocouples can operate at lower
temperatures than the Pt-base ones. Table1 reports the approximate compositions for positive and negative thermoelements of the letter designated thermocouples.
Figure5: Voltage vs Temperature relationship for letter-designated thermocouples |
|
||||||||||||||||||||||||
Table1: Approximate composition for thermoelements of letter-designated thermocouples |
All the voltage-temperature
relationships of the letter designated thermocouples are monotonic, but
not linear. For instance the type N thermocouple voltage output is
defined by the following 10 degree polynomials, where t is the
temperature in degree Celsius:
Equation2 |
The coefficients Ci are reported in Table2.
In order to have a linear voltage-temperature relationship the Seebeck coefficient should be constant with temperature (see Equation1); however the Seebeck coefficient is temperature dependent, as shown for instance for type K thermocouple in Figure6. Additional details on the voltage-temperature relatinships for letter designated thermocouple can be found at:
|
||||||||||||||||||||||||||||||||||||
Table2: Type N thermocouple coefficents |
Figure6: Type K Seebeck coefficient vs Temperature |
That sure is a lot to digest.